Skip to content

Public FDA record

Z-0144-2012: Bard Ventralex" ST Medium Circle with Strap Hernia Patch Diameter 6.4cm /2.5", Catalog Number: 5950008.

Recall record Z-0144-2012 for Davol, Inc., Sub. C. R. Bard, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0144-2012
Company in FDA record
Davol, Inc., Sub. C. R. Bard, Inc.
FDA classification date
Nov 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Mislabeled: Bard Ventralex" ST Medium Circle was packaged in a pre-printed carton identified for a Bard Ventralex" ST Large Circle.

Source facts

Action
Davol Inc. issued an Urgent; Medical Device Recall letter on 10/4/11 for the mislabeled Bard Ventralex ST Medium Circle. The letter identified the affected product and explained the reason for recall. Customers are to immediately examine their inventory and quarantine any affected product. In addition, if any product was further distributed then those customers should be notified of the recall as well. Accounts are to contact Davol Customer Service at 1-800-556-6275 for instructions on how to return product to Davol. The attached recall effectiveness check form should be completed and returned. Questions should be directed towards Davol Customer Service Department at 1-800-556-6275 or C. R. Bard's Medical Services & Support Department at 1-800-562-0027.
Initiated
2011-10-04
Root cause
Packaging process control
Terminated
2013-09-20
Distribution
Nationwide Distribution
Recall number
Z-0144-2012
Recall status
Terminated
CDRH record ID
104352
FDA classified
2011-11-01
Recall event number
60057

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗