- FDA identifier
- Z-0144-2015
- Company in FDA record
- IMRIS Inc
- FDA classification date
- Oct 28, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One of the three pins that is used to hold the ORT200 table alignment block in place on the floor of the MR Suite went missing during testing at a health care facility.
Source facts
- Action
- Consignees were sent on 10/01/2014 an IMRIS "Urgent Field Safety Notice" letter dated September 25, 2014. The letter described the problem and the product involved in the recall. The letter informed consignees that the pins will be replaced with pins made of non-ferrous material. Customers are to ensure that users are aware of the changes and the existing procedure for table alignment and managing ferrous content in the magnet room. Contact customer support at 1-866-475-0525 for questions.
- Initiated
- 2014-10-01
- Root cause
- Use error
- Terminated
- 2015-02-20
- Distribution
- Worldwide Distribution -- USA, including the states of CA, FL, NH, NC, TN, and UT; and, the countries of AUSTRALIA, CANADA, and JAPAN.
- Recall number
- Z-0144-2015
- Recall status
- Terminated
- CDRH record ID
- 130127
- FDA classified
- 2014-10-28
- Recall event number
- 69265