Skip to content

Public FDA record

Z-0145-2022: Herniatome, 17G x 15cm, Lumbar Type, Model Number 5091745

Recall record Z-0145-2022 for ADRIA SRL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0145-2022
Company in FDA record
ADRIA SRL
FDA record creation date
Oct 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recalling firm was notified by the contract sterilizer that the process parameters of the sterilization cycles were not followed.

Source facts

Action
An Urgent Field Safety Notice dated 5/17/21 was sent to customers. Actions required to distributors and economic operators 1. Immediately suspend deliveries, identify and quarantine all the devices in your possession which are listed in Annex no. 1 "List of Impacted batches". 2. Share this Field Safety Notice within your organization with all interested parties. If you have distributed the products covered by this FSN to third parties, identify these subjects and forward this letter to them immediately, communicating to each hospital the detailed list of the goods subject to this action that have been supplied by you, using the template of the Annex no. 2 "Letter of the Distributor to Hospitals" Table A1 (making sure to fill in Table A1 with the detail of the article codes and the lots destined for that hospital). 3.Fill in and sign the attached Annex no. 3 "Acknowledgment of the distributor" specifying the quantity of the quarantined goods, including their lot number, their code and inform Adria Srl by sending an email to qa@adriamedical.com and xport@adriamedical.com as soon as possible / within 10 days of receipt of this letter. 4.Adria Srl will contact you to organize the collection of the goods. Adria Srl will replace the goods as soon as possible. Actions required of hospitals and healthcare facilities 1. Identify and immediately stop using the concerned products. Place in quarantine all items listed in Table A1 that are still present at your facilities. 2.Fill in and sign the attached Annex no. 4 Acknowledgement of hospitals and health facilities specifying the quantity of the goods placed in quarantine, including their lot number, their code and inform your dealer and Adria Srl by sending an email to qa@adriamedical.com and export@adriamedical.com as soon as possible and in any case within 10 days of receipt of this letter. 3. Wait for information from Adria Srl for the actions regarding the concerned devices. Corrective Actions in place Adria Srl ha
Initiated
2021-05-17
Root cause
Other
Terminated
2023-11-06
Distribution
The products were distributed to the following US states: NJ and NY.
Recall number
Z-0145-2022
Recall status
Terminated
CDRH record ID
188959
FDA record created
2021-10-19
Recall event number
88543

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗