Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.
Source facts
Action
On September 12, 2025, Kico Knee Innovation Company issued a recall notice to affected consignees. Kico asked consignees to take the following actions:
1. No actions are required by the surgeon users.
2. For the plans affected, Kico has quarantined these as they are not suitable for clinical use.
3. A long-term solution has been implemented to return the system to the specification of the most recently cleared device. We will contact surgeon users to login and re-plan their cases for upcoming surgeries.
Initiated
2025-09-19
Root cause
Software change control
Distribution
US Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY, CA, WI, MI, PA, WA, CO, AL.
Recall number
Z-0145-2026
Recall status
Open, Classified
CDRH record ID
216120
FDA classified
2025-10-14
Recall event number
97618