Open original FDA record ↗Recall record
- FDA identifier
- Z-0145-2027
- Company in FDA record
- Ferno-Washington Inc
- FDA classification date
- Oct 9, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.
Source facts
- Action
- Firm began notifying customers on April 28, 2026. Customers were informed that their Power F2 system would be receiving a firmware update. PX2 firmware was also updated to ensure consistent communication with PF2 Connect. Instruction to Customers: Until your Power F2 system has been updated: - continue to use the device in accordance with the instructions for use - if powered functionality is not available, transition to manual operation as needed -ensure personnel are familiar with manual operations procedures
- Initiated
- 2026-04-28
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, ID, IL, KY, LA, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Netherlands and Poland.
- Recall number
- Z-0145-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222747
- FDA classified
- 2026-10-09
- Recall event number
- 99861