Skip to content

Public FDA record

Z-0146-2023: Fogarty Arterial Embolectomy Catheters Model Number: 120602FP 120403FP 120803FP 120404FP 120804FP 120805

Recall record Z-0146-2023 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0146-2023
Company in FDA record
Edwards Lifesciences, LLC
FDA classification date
Oct 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.

Source facts

Action
On 10/06/2022, Edwards Lifesciences sent an "URGENT MEDICAL DEVICE RECALL" Letter via FedEx 2nd Day to customer informing them that certain lots of Fogarty arterial embolectomy catheters and Fogarty biliary balloon probes are exhibiting latex deterioration in the balloon. It was determined that storage of these devices in the same room room as high energy ionizing radiation sources which can act as an ozone generator to the immediate environment, can accelerate latex deterioration and subsequent balloon failures. Advice on action to be taken by users: It is requested that they do not store Fogarty arterial embolectomy catheters and Fogarty biliary balloon probes in rooms with high energy ionizing radiation sources that could generate ozone (e.g. Fluoroscopy machines, X-ray machines, UV lights, HVAC sanitation equipment, etc). Post the attached notification as a constant reminder to their staff of the issue. - If they have devices that have been exposed to the equipment described above, please return the affected inventory with the model and lot number referenced above. - If they have devices that have not been exposed to the equipment above (e.g. stored away from high energy ionizing radiation source), they may continue to use the devices per the IFU. Review their inventory and quarantine any exposed product until prepared for return to Edwards Lifesciences. Follow the instructions included in the enclosed customer acknowledgement form and return within 5 days of receipt of this notification. Edwards Customer Service 1-800-424-3278 from 5:00 a.m. to 4:30 p.m. PST.
Initiated
2022-10-06
Root cause
Packaging
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Albania, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bulgaria, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guam, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Malta, Mauritius, Morocco, Netherlands, Nicaragua, Northern Ireland, Oman, Panama, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates (UAE), United Kingdom.
Recall number
Z-0146-2023
Recall status
Open, Classified
CDRH record ID
196175
FDA classified
2022-10-28
Recall event number
90905

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗