Skip to content

Public FDA record

Z-0146-2027: iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T

Recall record Z-0146-2027 for SI-BONE, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0146-2027
Company in FDA record
SI-BONE, Inc.
FDA classification date
Oct 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.

Source facts

Action
On August 31, 2026 and September 2, 2026, SI-Bone, Inc. Chief Medical Officer spoke with affected surgeons via phone concerning the recall.
Initiated
2026-08-21
Root cause
Packaging
Distribution
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Recall number
Z-0146-2027
Recall status
Open, Classified
CDRH record ID
222762
FDA classified
2026-10-09
Recall event number
99868

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗