Skip to content

Public FDA record

Z-0147-2027: FORWARD Gunshot Wound Kit; Model: AFT-GSW;

Recall record Z-0147-2027 for INTEGRATED MEDCRAFT LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0147-2027
Company in FDA record
INTEGRATED MEDCRAFT LLC
FDA classification date
Oct 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.

Source facts

Action
On September 25, 2026 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Immediately discontinue use of any product matching the affected item(s) and lot number(s) listed in the item table above. Inspect inventory for affected product. If you have no products matching the affected item(s), please fill out the acknowledge receipt form (referenced below) and indicate that you have none of these products in inventory. 2. A copy of the Urgent Medical Device Recall from Safeguard Medical has been included for reference. 3. If you have product affected by this notice, fill out the attached Acknowledgement and Receipt Form and return it to quality@integratedmc.com. To ensure timely product replacement, please respond within 30 days from date of this notice. 4. If you do not have the recalled product in stock, please check the appropriate box on the form. 5. If you have further distributed any of the items referenced in this notification, provide your customers with a copy of this notification and request that they return the affected product directly to you or directly to Integrated MedCraft.
Initiated
2026-09-25
Root cause
Component change control
Distribution
US Nationwide distribution in the states of GA, VA.
Recall number
Z-0147-2027
Recall status
Open, Classified
CDRH record ID
222960
FDA classified
2026-10-09
Recall event number
99948

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.