Open original FDA record ↗Recall record
- FDA identifier
- Z-0148-2007
- Company in FDA record
- Advanced Medical Optics, Inc.
- FDA classification date
- Dec 14, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The lenses are incorrectly labeled as 11 diopter lenses; these lenses are actually 22.5 diopters.
Source facts
- Action
- On November 13, 2006, an 'URGENT MEDICAL DEVICE RECALL' letter was mailed to all affected US customers communicating their recall strategy regarding the removal of five (5) ZA9003 lenses in US distribution. This letter included in this correspondence is a listing of the serial numbers and the quantities of intraocular lenses shipped. This listing will act as your Facsimile Form to be returned to AMO for accountability of the lenses shipped to your facility. AMO will ensure the effectiveness of the recall by using a return facsimile form to capture the status of each of the lenses.
- Initiated
- 2006-11-29
- Root cause
- Other
- Terminated
- 2011-05-24
- Distribution
- Worldwide, including USA (CT, TX, NY), Canada, and UK.
- Recall number
- Z-0148-2007
- Recall status
- Terminated
- CDRH record ID
- 49638
- FDA classified
- 2006-12-14
- Recall event number
- 36943