GE Healthcare has become aware of missing labels on the alignment lasers.
Source facts
Action
GE Healthcare will execute a field action in which GE Healthcare Field Engineers will perform an inspection of all suspect MR systems. Any systems lacking any of the 3 laser warning labels will be remediated by applying the missing labels). We expect this field action will be completed by June 30, 2012.
1. Notification of all distributors and purchasers be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. The FDA district office coordinator noted below is to be included in the notifications.
2. Corrections be made at no cost to the purchasers.
Initiated
2011-09-15
Root cause
Labeling design
Terminated
2012-06-29
Distribution
Nationwide Distribution
Recall number
Z-0148-2012
Recall status
Terminated
CDRH record ID
104146
FDA classified
2012-02-08
Recall event number
59982