Skip to content

Public FDA record

Z-0148-2020: Siemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000 Product Usage: automated, c

Recall record Z-0148-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0148-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Oct 16, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below

Source facts

Action
Siemens issued an Urgent medical Device Correction, (UMDC ASI19-02.A.US) via Fed Ex and Urgent Field Safety Notification (UFSN ASI19-02.A.OUS) to all affected Customers beginning on 2019-07-30. The UFSN/UMDC explain the behaviors that can occur and instructs customers on actions to be taken for the issues described in the communication. Customers are requested to return the attached effectiveness check form via fax or email within 30 days. Field Service personnel will be sent a Service Knowledge Base describing the issue and instructing them how to deal with customer questions.
Initiated
2019-07-30
Root cause
Software design
Terminated
2022-06-10
Distribution
Worldwide Distribution - US Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Brazil Bulgaria Canada Chile Colombia Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Hungary India Indonesia Iran Ireland Israel Italy Japan Kuwait Latvia Malaysia Mexico Netherlands Nigeria Norway P.R. China Pakistan Philippines Poland Portugal Qatar Republic Korea Romania Russian Fed. Saudi Arabia Singapore Slovakia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
Recall number
Z-0148-2020
Recall status
Terminated
CDRH record ID
175626
FDA record created
2019-10-16
Recall event number
83529

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗