Open original FDA record ↗Recall record
- FDA identifier
- Z-0150-2008
- Company in FDA record
- Datascope Corporation
- FDA classification date
- Jan 17, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Display problems: If an ECG cable, which has been damaged due to the ingress of liquid or by mechanical trauma, is utilized with the Telepack, it may cause the Telepack to switch between the 3-lead and 5 lead-input modes. During the switching process, the digital heart rate displayed at the Panorama Central Station will be frozen, and subsquent arrthymia alarms will not be announced.
Source facts
- Action
- Urgent Product Field Correction notices were sent to hospital administrators on October 9, 2007, by certified mail, return receipt requested. The letter describes the scenario that causes the problem and its remediation It also describes how the software upgrade will fix the problem and alerts hospitals that a Datascope Service Representative will be contacting them to arrange for the upgrade.
- Initiated
- 2007-10-09
- Root cause
- Software design
- Terminated
- 2008-11-04
- Distribution
- Nationwide
- Recall number
- Z-0150-2008
- Recall status
- Terminated
- CDRH record ID
- 65387
- FDA classified
- 2008-01-17
- Recall event number
- 45405