Z-0150-2010: ABL 80 CO-OX SYSTEM, software versions 1.30 and 1.31 Additional Product Codes: CEM, CGA, CGZ, GHS, GKR,

Recall record Z-0150-2010 for Sendx Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0150-2010
Company in FDA record
Sendx Medical Inc
FDA classification date
Nov 10, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Inverness have recently become aware of a potential error contained in V1.30 and V1.31 software for the ABL 80 CO-OX platform. The error can occur if the analyzer is set to automatically upload patient records to Radiance or an HIS/LIS system. This can result in the reporting of incorrect results for one or more CO-OX derived parameters. Incorrect results can be displayed on the screen and the

Source facts

Action
Radiometer sent out a letter to consignees informing them of public reason for recall, and telling them that they have provided a copy of the upgrade software and instructions to correct this potential problem immediately. Software upgrade instructions were included with the recall letter. Consignees were informed that if they preferred that their local service representative perform the upgrade, to follow the Instructions for disabling auto upload of patients records. SOFTWARE UPGRADE INSTRUCTIONS: The upgrade software must be installed via the Software Upgrade option under Menu>Utilities>Software Upgrade. It is not a "boo table" CD. ~The Upgrade process will take approximately 10 minutes. The analyzer should be allowed to complete the entire process without user interaction. Do not attempt to turn the analyzer on or off manually at any time during the process. ~The upgrade process will include the following general steps: 1. Automatic power down and reboot of the analyzer 2. File transfer process which includes a variety of progress bars to communicate the status 3. Automatic power down and reboot of the analyzer a second time 4. A dialog box on the main menu screen that states, "ABL80 FLEX software has been upgraded to version 1.32." This signifies a successful upgrade process. ~If a solution pack and a sensor cassette are already installed on the analyzer, their installation status will not be affected by the software upgrade. ~All stored data will be retained in all logs They were also asked to complete and return the attached fax form, with their signature, to 1-800-736-0601. If their were any questions regarding the letter, or upgrade, to contact Techline at 1-800-736-0600 opt.
Initiated
2009-02-20
Root cause
Other
Terminated
2012-07-16
Distribution
Domestically and Worldwide: Germany, France, Spain, United Kingdom, Poland, Austria, Italy, Tokyo, Australia, and Denmark Domestically to OH
Recall number
Z-0150-2010
Recall status
Terminated
CDRH record ID
84010
FDA classified
2009-11-10
Recall event number
52791

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗