Skip to content

Public FDA record

Z-0150-2024: Monoject" 60 mL Syringe Luer-Lock Tip Soft Pack

Recall record Z-0150-2024 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0150-2024
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Oct 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Demonstrated recognition and compatibility issues with syringe infusion pumps.

Source facts

Action
Cardinal Health notified customers on about 09/20/2023 via "URGENT MEDICAL DEVICE PRODUCT CORRECTION" letter sent FedEx next day delivery. Customers were instructed to: 1. REVIEW your inventory for the affected product codes and lots. Location of product code and lot are shown in above listed table and below example labeling. 2. COMMUNICATE with all personnel that utilize the Cardinal Health Monoject" Luer-Lock Tip syringes (1, 6, 12, 20, 35 and 60 mL) that they should not be used with syringe infusion pumps. 3. POST a copy of this notification in your storeroom where the product is stored. 4. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this medical device product correction and share a copy of this notice. 5. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB FieldCorrectiveAction@cardinalhealth.com, whether you have affected product or not. complete and return the provided acknowledgement form. For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Initiated
2023-09-20
Root cause
Vendor change control
Distribution
US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Recall number
Z-0150-2024
Recall status
Open, Classified
CDRH record ID
203389
FDA classified
2023-10-26
Recall event number
93075

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗