Z-0151-2007: VITROS Chemistry Products GLU DT Slides GEN 56, REF (Catalog #) 153 2316, 25 slides per box. VITROS Chemi

Recall record Z-0151-2007 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0151-2007
Company in FDA record
Ortho-Clinical Diagnostics
FDA classification date
Nov 4, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Positively biased results when using the VITROS Chemistry Products GLU DT Slides GEN 56.

Source facts

Action
On 9/26/06, the recalling firm notified its foreign affiliate consignees of this 'field correction', and instructed them to inform their customers who received these slides of this issue, provided them with an updated Calibration Data Module and instructed to re-calibrate these slides for continued use. On 9/27/06, all US end-use consignees were informed of this issue via Federal Express overnight letter (dated 9/27/06), and instructing them to re-calibrate these slides for continued use after they have installed the Calibration Data Module. The new CDMs were sent to all customers with VITROS DT 60/DT60 II Chemistry systems starting 9/27/06 concurrent with, but separate, from the correction letters. This includes consignees who purchased their GLU DT slides from distributors. On 9/28/06, all distributors were notified of this 'field correction', providing instructions and 'Questions & Answers' for them, and instructing them to inform their customers who received these slides of this issue. Calibration Data Modules were sent to all customers with VITROS DT 60/DT60 II Chemistry Systems.
Initiated
2006-09-27
Root cause
Other
Terminated
2012-08-03
Distribution
Worldwide Distribution
Recall number
Z-0151-2007
Recall status
Terminated
CDRH record ID
48778
FDA classified
2006-11-04
Recall event number
36509

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗