Open original FDA record ↗Recall record
- FDA identifier
- Z-0151-2008
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Feb 22, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use of an under-rated fuse will result in failure of the fuse and a loss of power to the vision cart, including the loss of power to any ancillary equipment that m
Source facts
- Action
- A product recall was initiated on October 4, 2007. Direct and Distributor consignees were sent an Important Product Notice, via electronic mail on October 5, 2007. The firm plans on sending a Field Service Engineer to inspect the system and make any necessary corrections.
- Initiated
- 2007-10-04
- Root cause
- Mixed-up of materials/components
- Terminated
- 2011-03-31
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, LA, MA, MD, MN, NC, NJ, NV, OH, PA, PR, SD, TN, TX, WA, and WI, and countries of Belgium, Ireland, Italy, Korea, Russia, and the United Kingdom.
- Recall number
- Z-0151-2008
- Recall status
- Terminated
- CDRH record ID
- 65719
- FDA classified
- 2008-02-22
- Recall event number
- 45518