Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).
Source facts
Action
Hamilton Medical issued a letter dated April 22, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that their Hamilton Medical Account Manager would contact them to coordinate the deactivation of the Ventilation Suppression feature. Customers were asked to review the Medical Device Field Safety Corrective Action.
Customers with questions were instructed to call 800-426-6331, Ext 215.
Initiated
2014-04-22
Root cause
Device Design
Terminated
2017-08-24
Distribution
Nationwide Distribution
Recall number
Z-0151-2016
Recall status
Terminated
CDRH record ID
140598
FDA classified
2015-11-04
Recall event number
72319