Reason stated in the source
The recall was initiated because Ameditech, Inc. received a complaint from US Diagnostics (USD) that First Check Drug Screen Cups were packaged in the ProScreen Cup boxes. Some of US Diagnostics' customers received ProScreen 5-panel cups that contain First Check THC cups. The recalled products are mislabeled. If the USD customer receives the FirstCheck tes, there will be no instructions for use
Source facts
- Action
- The recall communication was initiated on June 9, 2011 with Ameditech Inc. forwarding an Urgent Medical Device Recall notification with an attached Verification Form (via phone and written notification) to the affected distributor. The letter provides the distributor with an explanation of the problem identified and actions to be taken. The distributor was required to: cease distribution of the affected lot and provide remaining inventory count to Ameditech; destroy any remaining inventory of these lots and document destruction per site requirements; and ensure that all user of the device have received a copy of the Urgent Medical Device Recall. In addition, the distributor was instructed to complete and return the enclosed Verification Form with in 10 days. For any questions about the information contained in this notice, contact Ameditech at amd.complaint@alere.com or call 858-535-1968.
- Initiated
- 2011-06-07
- Root cause
- Packaging process control
- Terminated
- 2012-01-18
- Distribution
- Nationwide Distribution (USA) including the state of Alabama.
- Recall number
- Z-0152-2012
- Recall status
- Terminated
- CDRH record ID
- 104433
- FDA classified
- 2011-11-03
- Recall event number
- 60100