Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Source facts
Action
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Initiated
2025-09-09
Root cause
Package design/selection
Distribution
US Domestic distribution to Texas and Michigan.
Recall number
Z-0152-2026
Recall status
Open, Classified
CDRH record ID
216096
FDA classified
2025-10-14
Recall event number
97605