Z-0153-2023: CritiCool, thermoregulation device

Recall record Z-0153-2023 for Belmont Instrument Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0153-2023
Company in FDA record
Belmont Instrument Corporation
FDA classification date
Oct 31, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may potentially lead to delayed treatment and a failure to deliver the prescribed thermoregulation therapy. To date, the firm has received one report of death potentially related to this issue.

Source facts

Action
On September 28, 2022, the firm issued Urgent Medical Device Recall letters to affected customers. If the device is operating properly, customers may continue to use the device as intended with increased oversight during use while awaiting inspection and repair. The customer was provided with actions to take if a HALT 4 message is displayed on the device. While customers wait for their devices to be inspected and repaired, Belmont Medical Technologies recommends that you evaluate your clinical needs and consider planning for alternative methods of thermoregulation in case of failure. Belmont Medical Technologies will dispatch service technicians to customer sites to inspect and repair affected CritiCool devices. Customers will be contacted by the service technician to schedule a convenient time for the inspection and repair of customer devices. For any questions regarding this action, technical assistance or to report an adverse event or product complaint, please contact Belmont Medical Technologies using the following information: 1-855-387-4547 TECHSUPPORT@BELMONTMEDTECH.COM
Initiated
2021-07-20
Root cause
Component design/selection
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Recall number
Z-0153-2023
Recall status
Open, Classified
CDRH record ID
196238
FDA classified
2022-10-31
Recall event number
90939

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗