Open original FDA record ↗Recall record
- FDA identifier
- Z-0153-2023
- Company in FDA record
- Belmont Instrument Corporation
- FDA classification date
- Oct 31, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may potentially lead to delayed treatment and a failure to deliver the prescribed thermoregulation therapy. To date, the firm has received one report of death potentially related to this issue.
Source facts
- Action
- On September 28, 2022, the firm issued Urgent Medical Device Recall letters to affected customers. If the device is operating properly, customers may continue to use the device as intended with increased oversight during use while awaiting inspection and repair. The customer was provided with actions to take if a HALT 4 message is displayed on the device. While customers wait for their devices to be inspected and repaired, Belmont Medical Technologies recommends that you evaluate your clinical needs and consider planning for alternative methods of thermoregulation in case of failure. Belmont Medical Technologies will dispatch service technicians to customer sites to inspect and repair affected CritiCool devices. Customers will be contacted by the service technician to schedule a convenient time for the inspection and repair of customer devices. For any questions regarding this action, technical assistance or to report an adverse event or product complaint, please contact Belmont Medical Technologies using the following information: 1-855-387-4547 TECHSUPPORT@BELMONTMEDTECH.COM
- Initiated
- 2021-07-20
- Root cause
- Component design/selection
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Recall number
- Z-0153-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196238
- FDA classified
- 2022-10-31
- Recall event number
- 90939