Open original FDA record ↗Recall record
- FDA identifier
- Z-0154-06
- Company in FDA record
- Medrad Inc
- FDA classification date
- Nov 16, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
the heat maintainer, an accessory, may overheat and deform the disposable syringe
Source facts
- Action
- The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/12/05 via certified mail. The notice provided instructions to apply warning labels and disconnect the heat maintainer until a field upgrade kit is available. In case where the heat maintainer is a necessity for specific procedures, the customer is being instructed to observe certain precautions.
- Initiated
- 2005-10-10
- Root cause
- Other
- Terminated
- 2007-08-15
- Distribution
- The products were shipped to medical facilities nationwide. The products were also shipped to Canada, Isreal, Italy, Denmark, Netherlands, Norway, Great Britain, France, Poland, Germany, Spain, Finland, Belgium, Sweden, Russia, Austraila, Cuba, Mexico, Peru, Hong Kong, Korea, Mylasia, Thialand, Taiwan, Singapore, India, China, Turkey, Saudia Arabia, and Japan
- Recall number
- Z-0154-06
- Recall status
- Terminated
- CDRH record ID
- 42540
- FDA classified
- 2005-11-16
- Recall event number
- 33846