Open original FDA record ↗Recall record
- FDA identifier
- Z-0154-2016
- Company in FDA record
- Berchtold Corp.
- FDA record creation date
- Oct 22, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the use of the TELETOM Series 4 & 6 there is potential for the controls and brakes not to work properly due to fluid intrusion within the control switches and/or the control board assembly.
Source facts
- Action
- Stryker sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. A Stryker Representative would visit each account for product inspection. During the inspection the Stryker Representative will remove the control switches and replace them with new control switches. Customers were instructed if any malfunction prior to inspection were noticed customers should call local Technical Support immediately at 800-243-5135.
- Initiated
- 2015-09-23
- Root cause
- Device Design
- Terminated
- 2016-11-17
- Distribution
- Worldwide Distribution - US (nationwide), Canada, and Germany
- Recall number
- Z-0154-2016
- Recall status
- Terminated
- CDRH record ID
- 140448
- FDA record created
- 2015-10-22
- Recall event number
- 72273