Skip to content

Public FDA record

Z-0154-2023: 18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL P

Recall record Z-0154-2023 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0154-2023
Company in FDA record
Insulet Corporation
FDA classification date
Nov 15, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Source facts

Action
On October 17, 2022, the firm notified customers via Urgent Medical Device Correction letters sent via email. The firm followed up with physical mailers. Customers will receive updated Omnipod DASH PDMs in the coming months. Customers do not need to call to receive an updated device, one will automatically be shipped to you when available. Customers were provided with additional precautions to follow until they receive replacement units. If you have any questions regarding the information provided in this Medical Device Correction, please visit our website at www.omnipod.com/insulet-alerts or call 1-800-641-2049 to speak with a dedicated staff member of the Insulet Customer Care team who will be available 24 hours a day, 7 days a week.
Initiated
2022-10-17
Root cause
Device Design
Terminated
2024-09-17
Distribution
Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.
Recall number
Z-0154-2023
Recall status
Terminated
CDRH record ID
196402
FDA classified
2022-11-15
Recall event number
90987

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗