Z-0155-2024: SOZO Bilateral Arm L-Dex Software

Recall record Z-0155-2024 for Impedimed Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0155-2024
Company in FDA record
Impedimed Limited
FDA classification date
Oct 24, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bilateral L-Dex assessment software does not have the same level of sensitivity to help detect early signs of lymphedema as the unilateral arm L-Dex assessment, which could result in under-recognition of early lymphedema, which could result in delay in early intervention, and more aggressive intervention.

Source facts

Action
On 8/17/23, correction notices were emailed to customers. Customers were asked to take the following actions: We recommend that you confirm the risk profile of your bilateral arm patients in MySOZO and take the following actions: - For patients at unilateral risk for lymphedema (over 90% of breast cancer patients), confirm that the unilateral arm L-Dex assessment is selected in the patient profile and continue to screen for early signs of lymphedema. - For patients at bilateral risk for lymphedema (3-10% of breast cancer patients), guidelines suggest routine screening using clinical exam and symptom assessment. - Complete and return the response form via link: https://forms.office.com/r/kafiuDbw21 In September, the firm will be issuing a SOZO software update to make unilateral L-Dex the sole assessment option for arms. L-Dex assessments for legs will not be impacted. This correction letter should be shared with anyone within your organization who needs to be aware of the correction or with any organization where the device may have been transferred. If you have questions or need to report quality issues, call 1/877/247-0111, Option 4 (Mon-Fri, 8:00 am - 8:00 om ET) or email impedimed156@impedimed.com
Initiated
2023-08-17
Root cause
Device Design
Distribution
US Nationwide distribution in the states of IA, CA, IL, AK, NY, FL, KS, WI, MI, NC, GA, SD, KY, TX, MA, LA, AZ, RI, AR, CO, WY, OH, IN, MO, NJ, MD, CT, WA, PA, UT, VA, NM, ME, MN, DC, TN, NE, OR, ND, WV.
Recall number
Z-0155-2024
Recall status
Open, Classified
CDRH record ID
202960
FDA classified
2023-10-24
Recall event number
92943

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗