Skip to content

Public FDA record

Z-0156-04: Midwest XGT High Speed Handpiece with Fiber Optics and Push Button Chuck, catalog #790000; Manufactured f

Recall record Z-0156-04 for DentSply -Professional Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0156-04
Company in FDA record
DentSply -Professional Division
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Small cracks in the dental handpiece's chuck could allow the dental burs to drift out of the chuck over time.

Source facts

Action
Dealers were telephoned on 10/14-16/03 and sent follow-up faxed letters on the same dates. The dealers were informed of the potential for cracked bur chucking mechanisms in the handpieces, requested to examine their stocks for the affected serial numbers, using the faxed list of serial numbers shipped to each account, and return the affected handpieces to DENTSPLY via FedEx. The dealers were also requested to verify the serial number list to those handpieces shipped to customers. If the affected serial numbers were distributed, the customer was requested to supply a list of those customers with contact information to DENTSPLY. As those lists are received by DENTSPLY, end user letters are sent to those customers, informing them of the problem with the bur chucking mechanism and requesting the retutrn of the affected serial numbers for examination and replacement of the bur chucking mechanism if necessary.
Initiated
2003-10-14
Root cause
Other
Terminated
2004-08-03
Distribution
Nationwide, Canada and Mexico
Recall number
Z-0156-04
Recall status
Terminated
CDRH record ID
30313
FDA classified
2004-07-20
Recall event number
27578

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗