An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.
Source facts
Action
Omega Medical Imaging issued a Field Safety Advisory Notice to its consignees on 02/28/2023 by email. The notice explained the issue and pledged to evaluate all potentially impacted devices and mitigate the issue to bring the boom systems up to the latest safety design and standard of the newer models approved for field use.
Initiated
2023-02-28
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the country of Pakistan.
Recall number
Z-0156-2024
Recall status
Open, Classified
CDRH record ID
203289
FDA classified
2023-10-24
Recall event number
93018