Open original FDA record ↗Recall record
- FDA identifier
- Z-0157-2015
- Company in FDA record
- Terumo Cardiovascular Systems Corp
- FDA classification date
- Oct 31, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line
Source facts
- Action
- Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
- Initiated
- 2011-03-01
- Root cause
- Device Design
- Terminated
- 2014-11-06
- Distribution
- Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
- Recall number
- Z-0157-2015
- Recall status
- Terminated
- CDRH record ID
- 98878
- FDA classified
- 2014-10-31
- Recall event number
- 58185