The JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM. And also, a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM.
Source facts
Action
Smith & Nephew, Inc. issued an Urgent Medical Device Recall Notice to its consignees on 09/27/2023 via email and overnight mail. The notice explain the issue, potential risk, and requested the following actions:
Inspect inventory, locate, and quarantine the affected devices.
Sales representatives, district offices, or distributors were directed to notify their customers
The firm is seeking return of unused products.
Initiated
2023-09-27
Root cause
Component change control
Distribution
Worldwide distribution - US Nationwide and the country of Japan.
Recall number
Z-0157-2024
Recall status
Open, Classified
CDRH record ID
203572
FDA classified
2023-10-24
Recall event number
93156
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.