Z-0159-06: Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication De

Recall record Z-0159-06 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0159-06
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Nov 18, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A 'Check Flange' alarm may occur even though the syringe has been properly loaded into the AS50 Infusion Pump. The false alarm may occur during the syringe loading process or while an infusion is running. Also, the occlusion detection and alarm may be delayed when the pumps are used with larger volume syringes at low flow rates.

Source facts

Action
Urgent Device Correction letters dated 5/5/05 were sent to the AS50 customers via first class mail to the attention of the Director of Nursing. The accounts were informed of the false 'Check Flange' alarms due to a combination of certain syringe clamp assemblies on a specific serial numner range of pumps and the dimensional attributes of the Monoject 35 cc syringes. The accounts were requested to check the serial numbers on their AS50 pumps against the listing of serial numbers in the letter and contact Baxter Medication Delivery Services at 1-800-843-7867 for information regarding the return process to screen and replace the syringe clamp assemblies. The accounts were also informed of the delay in the detection of a downstream occlusion when the AS50 pump is used with large volume syringes at low flow rates. To minimize the occlusion detection times at low flow rates, Baxter recommended that the customers take the following actions: * Use the smallest possible syringe size for the volume of fluid to be delivered; * Select the low occlusion pressure setting to minimize the detection time; * Utilize infusion sets with thick-walled, small bore tubing to minimize the detection time; and * Always verify that the clamping device or stopcock is open in the fluid path prior to starting an infusion. The accounts were requested to complete and fax back the enclosed reply form indicating that they have examined their inventory for AS50 pumps, and listed the affected serial numbers found to be returned to Baxter for correction. Baxter extended the recall on 1/24/06 to 43 customers who received the pumps after the "check Flange Alarm" issue had been corrected in production, but had not been notified of the "Delay in Detection of Downstream Occlusion" issue. Copies of the Operator's Manual Addendum addressing the delay issue were included with the letter. Baxter also sent copies of the Operator's Manual Addendum to all of the accounts who received the 5/5/05 le
Initiated
2005-05-06
Root cause
Other
Terminated
2008-06-23
Distribution
Nationwide including Puerto Rico, and internationally to Colombia, Chile, Brazil, Saudi Arabia and Canada.
Recall number
Z-0159-06
Recall status
Terminated
CDRH record ID
38881
FDA classified
2005-11-18
Recall event number
31992

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗