Skip to content

Public FDA record

Z-0159-2007: Jostra HL-20 Heart Lung Machine - guiding pins

Recall record Z-0159-2007 for MAQUET Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0159-2007
Company in FDA record
MAQUET Inc.
FDA classification date
Nov 7, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.

Source facts

Action
An 'Important Safety Notice - Please Read Immediately' letter was sent to all US customers on 10/3/06 by FedEx Return Receipt.
Initiated
2006-10-03
Root cause
Other
Terminated
2007-12-19
Distribution
Worldwide
Recall number
Z-0159-2007
Recall status
Terminated
CDRH record ID
48261
FDA classified
2006-11-07
Recall event number
36255

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗