Open original FDA record ↗Recall record
- FDA identifier
- Z-0159-2011
- Company in FDA record
- Impac Medical Systems Inc
- FDA classification date
- Oct 28, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mistreatment-- The database conversion utility used during the upgrade errantly changes the start gantry angle to zero. Any affected treatment fields, if not already treated, would start delivering from zero instead of the original prescribed value.
Source facts
- Action
- ELEKTA IMPAC SOFTWARE sent a USER NOTICE dated February 19, 2010, to all customers. The notice identified the product, the problem, and the action to be taken by the customers. Customers were urged to verify the start angle of the gantry against the treatment field definition prior to delivery. Review all treatment plans and charts with conformal arcs. If a treatment plan with an errant gantry start angle was identified, customers were to cease using that plan immediately. A workaround was to be re-imported into their current version of MOSAIQ. IMPAC Support was to follow-up with each site and execute a script to detect and correct all affected treatment plans. For questions, concerns, and requests for upgrade, customers were to email support@impac.com or call their support center.
- Initiated
- 2010-02-19
- Root cause
- Process change control
- Terminated
- 2011-03-28
- Distribution
- Worldwide Distribution -- USA, Australia, Belgium, France, Germany, Hong Kong, Ireland, Israel, and Italy
- Recall number
- Z-0159-2011
- Recall status
- Terminated
- CDRH record ID
- 91633
- FDA classified
- 2010-10-28
- Recall event number
- 55712