Open original FDA record ↗Recall record
- FDA identifier
- Z-0160-06
- Company in FDA record
- Datex Ohmeda, Inc. dba GE Healthcare
- FDA classification date
- Nov 18, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Phototherapy medical device may lose its position and can cause a safety hazard to the patient.
Source facts
- Action
- The firm notified consignees by letter on 08/05/2005 and advised of field correction of the devices in 2 phases. The letter is lot specific and provides for an interim workaround and includes an adhesive sticker and instructions for tagging the device with a warning statement containing restriction for use : 'This phototherapy light should ONLY be used on patients in closed incubators.' The letter further advises of firm''s plan to provide free, mandatory repair at a later date, pending hardware solution.
- Initiated
- 2005-08-05
- Root cause
- Other
- Terminated
- 2007-10-24
- Distribution
- Product was distributed to medical facilities in the US and international subsidiaries and distributors.
- Recall number
- Z-0160-06
- Recall status
- Terminated
- CDRH record ID
- 41078
- FDA classified
- 2005-11-18
- Recall event number
- 32920