Skip to content

Public FDA record

Z-0160-2022: REF 8420.15.010, Bone Screw/Vite - Ti6AI4V, Self Tapping/Autofilettante, Dia=6.5mm, h=20mm, STERILE R, 21

Recall record Z-0160-2022 for Limacorporate S.p.A. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0160-2022
Company in FDA record
Limacorporate S.p.A
FDA record creation date
Oct 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the length of bone screws identified on labeling may not correspond to the actual length of the screw included.

Source facts

Action
On October 1, 2021 , Lima Corporate issued an "Urgent: Medical Device Recall" notification to all affected consignees via email. In addition to informing consignees about the recall product, the firm asked consignees to take the following actions: 1. Check your stock to locate and quarantine the affected devices received. Affected devices are required to be sent back to LimaCorporate together with a hard copy of the attached Response Form; 2. Fill out, sign and send the attached Response Form to the email address pms@limacorporate.com, as a confirmation that You have read and acknowledged the content of this Field Safety Notice; 3. If the product was further distributed, please provide those customers with a copy of this customer notification and response form. Once completed, please forward the response form to pms@limacorporate.com. 4. If needed, please address any inquiry on this Field Safety Notice (FSN) to the email address medicalcomplaints@limacorporate.com. 5. This notice needs to be passed on all those who need to b. e aware within your organization. This Field Safety Notice will be sent to the Competent Authorities of the Countries involved in this Field Safety Corrective Action. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax
Initiated
2021-10-01
Root cause
Mixed-up of materials/components
Terminated
2024-06-27
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, FL, MO, KS, CO, TX, KS, IL, PA, NY, NC, MI, MA, OK and the countries of Austria, Croatia, Germany, Ireland, Italy, Japan, Korea, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden.
Recall number
Z-0160-2022
Recall status
Terminated
CDRH record ID
189607
FDA record created
2021-10-21
Recall event number
88812

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗