Skip to content

Public FDA record

Z-0160-2024: Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085

Recall record Z-0160-2024 for Omega Medical Imaging, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0160-2024
Company in FDA record
Omega Medical Imaging, Inc.
FDA classification date
Oct 24, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.

Source facts

Action
Omega Medical Imaging issued a Field Safety Advisory Notice to its consignees on 09/08/2023 via email. The notice explained the problem and the risk. The firm will inspect all affected devices to ensure the proper fasteners were installed on the unit.
Initiated
2023-09-08
Root cause
Device Design
Distribution
Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma
Recall number
Z-0160-2024
Recall status
Open, Classified
CDRH record ID
203492
FDA classified
2023-10-24
Recall event number
93123

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗