Open original FDA record ↗Recall record
- FDA identifier
- Z-0161-2009
- Company in FDA record
- MAQUET Inc.
- FDA classification date
- Oct 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During installation or cleaning there is a risk of crushing the finger by handling the head rest in the area of the gas strut while simultaneously actuating the release lever.
Source facts
- Action
- Urgent Device Correction letters were sent beginning September 26, 2008 via Fed Ex. Letters identified the potential for finger crush in conjunction with the head rest. Proper operating instructions are provided as well as a warning sticker to be applied to the head rest indicating proper positioning. All letters included a customer response form to be completed by each institution. Information requested is: that device correction letter has been read and instructions followed; that the corrected operating instructions have been received; that the warning sticker has been properly placed on the headrest, or that they do not have the head rest identified in the device correction letter. Questions should be addressed to USA Maquet Inc at 1-888-627-8383 #1.
- Initiated
- 2008-09-26
- Root cause
- Other
- Terminated
- 2009-04-16
- Distribution
- Nationwide Distribution
- Recall number
- Z-0161-2009
- Recall status
- Terminated
- CDRH record ID
- 73774
- FDA classified
- 2008-10-16
- Recall event number
- 49545