Open original FDA record ↗Recall record
- FDA identifier
- Z-0162-2015
- Company in FDA record
- Merit Medical Systems, Inc.
- FDA classification date
- Oct 31, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.
Source facts
- Action
- Consignees were notified via letter on September 26, 2014. Written notification included product identification information, instruction to immediately quarantine any devices and discontinue use, ensure all personnel to whom devices were distributed are made aware of this field action, and instructions to contact their Merit representative. An email notification was sent to US Merit sales representatives with accounts that had received affected product. The affected sales representatives were provided a Customer Response Form applicable to their account, a representative copy of communications sent directly to their affected accounts, and a copy of the Frequently Asked Questions. Merit sales representatives were instructed to visit the account and assist with the return of the affected units back to the Merit Field Assurance Department.
- Initiated
- 2014-09-26
- Root cause
- Mixed-up of materials/components
- Terminated
- 2014-11-19
- Distribution
- Nationwide and to India. No govt/military/VA consignees.
- Recall number
- Z-0162-2015
- Recall status
- Terminated
- CDRH record ID
- 130706
- FDA classified
- 2014-10-31
- Recall event number
- 69452