Skip to content

Public FDA record

Z-0162-2016: Signature Patient-Specific Surgical Guides; patient specific instruments used as a guide during surgical

Recall record Z-0162-2016 for Materialise N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0162-2016
Company in FDA record
Materialise N.V.
FDA classification date
Oct 23, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential inaccuracies in the alignment of the guides. Surgery was inaccurately planned because the patient image data was incorrectly processed.

Source facts

Action
On September 16, 2015 Materialise NV identified four cases where not all available image sets were used to design the case. On September 18, 2015, prior to the surgery dates Materialise informed (phone call) the Signature Services Manager of Zimmer Biomet that inaccuracies in the alignment of the knee might be observed with the use of the guides due to this issue. Biomet was asked to segregate the packages for these four cases and to return them to the Materialise US office. On September 23, 2015 this was followed up by e-mail notification letter with response forms. Biomet was requested to contact Materialise customer service to return these four cases and arrange a replacement set of patient specific surgical guides Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. E-mail : ortho-quality@materialise. be Customer Service Materialise USA LLC 44650 Helm Court Plymouth, Michigan 48170 734-259-7033
Initiated
2015-09-23
Root cause
Use error
Terminated
2016-02-12
Distribution
Distributed in MI, ID, TX, and in France.
Recall number
Z-0162-2016
Recall status
Terminated
CDRH record ID
141084
FDA classified
2015-10-23
Recall event number
72425

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗