Z-0162-2027: LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.

Recall record Z-0162-2027 for Magellan Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0162-2027
Company in FDA record
Magellan Diagnostics, Inc.
FDA classification date
Oct 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.

Source facts

Action
Magellan notified consignees on about September 4, 2026, via telephone call indicating a written notification of the recall would follow. On 09/17/2026, a notification letter was emailed to consignees with the following instructions: 1. Complete, sign, and return the CONFIRMATION OF NOTIFICATION FORM. 2. Review your inventory for the affected LeadCare II Blood Lead Test Kit lot identified. 3. Immediately discontinue use of any remaining affected kits. 4. Return any recalled product still in your possession to Magellan Diagnostics, Inc. for destruction.
Initiated
2026-09-04
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY.
Recall number
Z-0162-2027
Recall status
Open, Classified
CDRH record ID
222454
FDA classified
2026-10-15
Recall event number
99776

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗