Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.
Source facts
Action
Stryker Spine sent an Urgent Medical Device Field Correction Notification letter beginning July 5, 2012, to all affected customers. Teh letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check the drill bit length in each respective drill guide prior to surgery to ensure the correct length protrudes through the guide. Customers with questions should call 201-760-8298. Customers were asked to report any adverse events or product quality problems to 1-888-457-7463.
Initiated
2012-07-05
Root cause
Device Design
Terminated
2013-01-28
Distribution
Nationwide distribution.
Recall number
Z-0163-2013
Recall status
Terminated
CDRH record ID
113355
FDA classified
2012-10-29
Recall event number
63272