Z-0163-2015: Bard Dimension Stone Basket Product Usage: The Bard Dimension Stone basket is a teardrop-shaped baske

Recall record Z-0163-2015 for C.R. Bard, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0163-2015
Company in FDA record
C.R. Bard, Inc.
FDA classification date
Oct 31, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The product is non-sterile.

Source facts

Action
Bard sent an Urgent: Medical Device Recall letter to each customer on 10/7/2014 with proof of delivery notice. Wholesalers were asked to conduct a sub-recall. Letter identify the affected product, problem and actions to be taken. Customers were instructed to examine their inventory and quarantine product subject to recall, do not use or further distribute any of the affected product, remove and return affected product to Bard Medical Division, and if the product have been further distributed identify your customers and notify them at once of the product recall. Customers were instructed to complete the enclosed Recall & Effectiveness Check Form and fax to 1-770-784-6469. For questions call 1-770-784-6471.
Initiated
2014-10-07
Root cause
Release of Material/Component prior to receiving test results
Terminated
2015-06-12
Distribution
Worldwide Distribution in the states of AL, FL, NJ, NY, NC, PA, SC, TX, WI, and countries of Germany, Italy, Kuwait, and Luxembourg.
Recall number
Z-0163-2015
Recall status
Terminated
CDRH record ID
130716
FDA classified
2014-10-31
Recall event number
69462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗