Open original FDA record ↗Recall record
- FDA identifier
- Z-0163-2020
- Company in FDA record
- Deerfield Imaging, Inc.
- FDA record creation date
- Oct 18, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull with insufficient force; the clinical staff unknowingly pinning the patient's skull with excessive force; or the clinical staff identifying the issue and not being able to continue with the procedure.
Source facts
- Action
- The firm disseminated a field safety notice on 06/26/2019 by letter. The notice explained the problem and requested that the consignee inspect the skull clamp prior to use and be aware that overtightening of the drive knob may result in failure. The drive knobs will be replaced with an updated design when available.
- Initiated
- 2019-06-26
- Root cause
- Device Design
- Terminated
- 2020-10-29
- Distribution
- TN, NH, FL
- Recall number
- Z-0163-2020
- Recall status
- Terminated
- CDRH record ID
- 174540
- FDA record created
- 2019-10-18
- Recall event number
- 83278