Open original FDA record ↗Recall record
- FDA identifier
- Z-0164-2007
- Company in FDA record
- USA Instruments Incorporated
- FDA classification date
- Nov 9, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm determined that certain potential conditions for use of their medical device, outside of recommended practices, or operating manual descriptions, could result in a localized RF burn and/or electrical shock to a patient on which the device is being used.
Source facts
- Action
- The user/customers were notified of this recall with an ''Urgent Customer Safety Notification'' which was sent to the user/customers via Certified mail per recall letter dated 09/20/2006. The notification informs the customer(s) of this potential problem and the three (3) conditions under which this problem (patient becoming burned) could possibly occur. The notification goes on to inform the customers that the firm will be upgrading all 1.5T HD Head Neck & Spine Arrays so as to eliminate this potential hazard, at no cost to the customer and that a Service Representative will be contacting the customer in order to make arrangements to upgrade the coil on-site. A toll-free telelphone no. is provided in the notification letter for questions and/or comments. Lastly, the recall notification provides interim instructions which the customer should follow. The operator/technician should: ask that the patients place their hands at their sides during a scan; immediately discontinue the scan if the patient undergoing the scan starts to move; and stay in constant visual and oral contact with the patient undergoing the scan at all times. For the 3.0T HD Head Neck Spine Array, Model 2415545, for GE 3.0T Excite MR System, The user/customers were notified of this recall with an ''Urgent Customer Safety Notification'' which was sent to the user/customers via Certified mail per recall letter dated 10/13/2006.
- Initiated
- 2006-09-20
- Root cause
- Other
- Terminated
- 2012-04-13
- Distribution
- Worldwide, including USA, (states of CA, NJ, PA, MN, IA, MA, WI, WA, and FL) Switzerland, the United Kingdom, France, Canada and Finland.
- Recall number
- Z-0164-2007
- Recall status
- Terminated
- CDRH record ID
- 49150
- FDA classified
- 2006-11-09
- Recall event number
- 36612