Z-0164-2007: 3.0T HD Head Neck Spine Array, Model 2415545, for GE 3.0T Excite MR System.

Recall record Z-0164-2007 for USA Instruments Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0164-2007
Company in FDA record
USA Instruments Incorporated
FDA classification date
Nov 9, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm determined that certain potential conditions for use of their medical device, outside of recommended practices, or operating manual descriptions, could result in a localized RF burn and/or electrical shock to a patient on which the device is being used.

Source facts

Action
The user/customers were notified of this recall with an ''Urgent Customer Safety Notification'' which was sent to the user/customers via Certified mail per recall letter dated 09/20/2006. The notification informs the customer(s) of this potential problem and the three (3) conditions under which this problem (patient becoming burned) could possibly occur. The notification goes on to inform the customers that the firm will be upgrading all 1.5T HD Head Neck & Spine Arrays so as to eliminate this potential hazard, at no cost to the customer and that a Service Representative will be contacting the customer in order to make arrangements to upgrade the coil on-site. A toll-free telelphone no. is provided in the notification letter for questions and/or comments. Lastly, the recall notification provides interim instructions which the customer should follow. The operator/technician should: ask that the patients place their hands at their sides during a scan; immediately discontinue the scan if the patient undergoing the scan starts to move; and stay in constant visual and oral contact with the patient undergoing the scan at all times. For the 3.0T HD Head Neck Spine Array, Model 2415545, for GE 3.0T Excite MR System, The user/customers were notified of this recall with an ''Urgent Customer Safety Notification'' which was sent to the user/customers via Certified mail per recall letter dated 10/13/2006.
Initiated
2006-09-20
Root cause
Other
Terminated
2012-04-13
Distribution
Worldwide, including USA, (states of CA, NJ, PA, MN, IA, MA, WI, WA, and FL) Switzerland, the United Kingdom, France, Canada and Finland.
Recall number
Z-0164-2007
Recall status
Terminated
CDRH record ID
49150
FDA classified
2006-11-09
Recall event number
36612

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗