Open original FDA record ↗Recall record
- FDA identifier
- Z-0164-2018
- Company in FDA record
- Medline Industries Inc
- FDA record creation date
- Nov 21, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The locking mechanism on the belt may fail to secure/hold the strap when in locked position potentially allowing the belt strap to slip through the buckle.
Source facts
- Action
- Medline Industries sent an Immediate Action Required Recall Notification letter via First Class mail dated August 10, 2017. The recall was later expanded and customers were notified via letter of the additional lots dated August 21, 2017. Instructions in both letters included to check inventory and quarantine any affected product, complete and return the included response form, coordinate the return of all affected product, and notify customers if the product was further distributed. Additional customers were identified and notified of the recall via the same process on September 19th, 2017. For further questions, please call (866) 359-1704.
- Initiated
- 2017-08-10
- Root cause
- Device Design
- Terminated
- 2020-04-24
- Distribution
- USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NV, NY, OH , OK, OR, PA, RI, SC, TN, TX, VA, VT, WA, WI and WV.
- Recall number
- Z-0164-2018
- Recall status
- Terminated
- CDRH record ID
- 159346
- FDA record created
- 2017-11-21
- Recall event number
- 78363