Skip to content

Public FDA record

Z-0165-04: AEROSET Immunoglobulin M, list number 1E01-01, manufactured by Abbott Laboratories, South Pasadena, CA

Recall record Z-0165-04 for Abbott Laboratories Diagnostic Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0165-04
Company in FDA record
Abbott Laboratories Diagnostic Div
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Assay may produce falsely low IgM results without error codes when testing specimens contain elevated serum paraproteins associated with various myeloma or other malignancies.

Source facts

Action
Consignees were provided with assay parameters for the RCD flag which may enhance the detection of these type of specimens. The assay parameter settings for the RCD flag were distributed to all AEROSET consignees in the United States by direct mail and to country managers worldwide via a Product Information letter dated July 31, 2002. Also provided with the Product Information letter was an AEROSET IgM Application Sheet containing the updated assay parameters for the RCD flag.
Initiated
2002-07-31
Root cause
Other
Terminated
2003-11-26
Distribution
Nationwide to the following states: Texas, Illinois, Indiana, Tennessee, Alaska, Pennsylvania, New York, Florida, Massachusetts, Missouri, Alabama, North Carolina, California, Ohio, Utah, Michigan, Arizona, Wisconsin, Oklahoma, Maine, Kansas, Georgia, Mississippi, Connecticut, Rhode Island, Washington, Wyoming, Louisiana, Washington, Maryland, New Jersey, Arkansas, Iowa, South Carolina, Colorado, Minnesota, North Dakota, and West Virginia. Worldwide to the following countries: Argentina, Australia, Brazil, Germany, England, Canada, Chile, Colombia, Dominican Republic, Denmark, France, Hong Kong, Japan, Korea, Singapore, Guatemala, New Zealand, Thailand, Turkey, Spain, Italy, South Africa, Sweden, Greece, Portugal, Venezuela, Pakistan, Finland, Ireland, and Taiwan.
Recall number
Z-0165-04
Recall status
Terminated
CDRH record ID
29368
FDA classified
2004-07-20
Recall event number
27178

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗