Open original FDA record ↗Recall record
- FDA identifier
- Z-0165-2015
- Company in FDA record
- Zimmer Surgical Inc
- FDA classification date
- Oct 31, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Non-injury complaints from Europe identified a malfunction that made it possible for a handpiece to start by itself when the power source is connected.
Source facts
- Action
- Zimmer sent an Urgent Medical Device Correction letter via email to US distributors with traceable Fed Ex mailings to all consignees dated October, 2014. The letter identified the affected product, problem and actions to be taken. Customers were advised of the potential malfunction and the operating procedure for the handpiece per the Instructions for Use (IFU). Zimmer Surgical, as the importer, will provide notice to consignees in the US accordingly. In addition, units will receive an upgrade as part of the annual preventive maintenance per the Instructions for Use.
- Initiated
- 2014-10-06
- Root cause
- Device Design
- Terminated
- 2015-04-15
- Distribution
- US Nationwide Distribution in the states of: AK, AL,AR, AZ,CA,CO,CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA,ME, MI, MN, MO, NC, NE, NJ, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY.
- Recall number
- Z-0165-2015
- Recall status
- Terminated
- CDRH record ID
- 130512
- FDA classified
- 2014-10-31
- Recall event number
- 69424