Philips Medical Systems North America Co. Phillips
FDA classification date
Nov 14, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
Source facts
Action
On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
Initiated
2006-10-12
Root cause
Other
Terminated
2010-09-24
Distribution
Nationwide.
Recall number
Z-0166-2007
Recall status
Terminated
CDRH record ID
48925
FDA classified
2006-11-14