An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.
Source facts
Action
Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table.
For questions regarding this recall call 925-602-8083.
Initiated
2010-03-31
Root cause
Other
Terminated
2012-03-01
Distribution
Worldwide Distribution including USA
Recall number
Z-0166-2011
Recall status
Terminated
CDRH record ID
93284
FDA classified
2010-10-28
Recall event number
56352