Skip to content

Public FDA record

Z-0166-2012: Abbott brand IntraLase Femtosecond Laser System, Model 3; Catalog Numbers: 20004D/J/K, R20004J/K; Product

Recall record Z-0166-2012 for AMO Manufacturing USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0166-2012
Company in FDA record
AMO Manufacturing USA, LLC
FDA classification date
Nov 4, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System.

Source facts

Action
Abbott Medical Optics Inc. (AMO) issued an "ADVISORY NOTICE" letter dated October 12, 2011 to all affected customers. The letter identifies the affected products, the problem, and the actions needed to be taken. Customers were instructed to exercise caution when making cuts that approach the corneal endothelium. The letter includes a 1-page Operator's Manual addendum that clarifies the recommendations. The letter states that AMO will conduct a field correction to adjust the EOF on the affected FS Laser Systems. The letter includes a response form to be completed and returned as per the instructions. Customers were instructed to notify all appropriate staff/personnel of the Advisory Notice. For any further questions regarding the information or recommendations, contact an AMO representative at 1-800-266-3375
Initiated
2011-10-12
Root cause
Process design
Terminated
2013-05-01
Distribution
Worldwide Distribution - USA (nationwide)
Recall number
Z-0166-2012
Recall status
Terminated
CDRH record ID
104854
FDA classified
2011-11-04
Recall event number
60241

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗