- FDA identifier
- Z-0166-2013
- Company in FDA record
- Sorin Group USA, Inc.
- FDA classification date
- Oct 29, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.
Source facts
- Action
- The firm, Sorin Group USA, Inc., sent an "URGENT FIELD SAFETY NOTICE" dated September 20, 2012 to all affected customers by certified mail. The notice described the product, problem and actions to be taken. Customers were instructed to discontinue use of and segregate affected product for return and replacement to the firm; to complete and return the attached Customer Response Form via fax to: 303-467-6502 or email to: yvonne.feyerherm@sorin.com; and if they have transferred the affected products to a third party, to pass this information to the Sorin Group. Further questions may be directed to Sorin Group Customer Service at 1-800-650-2623.
- Initiated
- 2012-09-20
- Root cause
- Mixed-up of materials/components
- Terminated
- 2013-01-28
- Distribution
- Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.
- Recall number
- Z-0166-2013
- Recall status
- Terminated
- CDRH record ID
- 113635
- FDA classified
- 2012-10-29
- Recall event number
- 63381