Z-0166-2016: Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and dem

Recall record Z-0166-2016 for Perkin Elmer Life Sciences Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0166-2016
Company in FDA record
Perkin Elmer Life Sciences Inc
FDA record creation date
Oct 26, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It would be possible for an abnormal analytical test result value from an affected newborn to be reported with a default result interpretation of normal. The analytical test result value would be correct but the interpretation of the test result value would be incorrect.

Source facts

Action
Perkin Elmer issued the Recall Letter on 8/24/15 delivered to the customer by email .Software Services will be asked to make sure the customers receive the Recall Letter and implement the recommended corrective measures: Ensure all laboratory routine sample testing procedures are followed to avoid procedural anomalies. In the event a procedure anomaly occurs, review the final analytical result and the corresponding interpretation to confirm both are accurate before releasing the report. Once a PerkinElmer reporting error query is available, continue to run the query until the final corrective action is deployed to your site. The final corrective action will involve an installation update to the software. For further information, please contact your local PerkinElmer representative or specimen.gate.support@perkinelmer.com
Initiated
2015-08-24
Root cause
Software design
Terminated
2021-01-21
Distribution
US distribution to FL, GA, and NV; and Canada
Recall number
Z-0166-2016
Recall status
Terminated
CDRH record ID
139579
FDA record created
2015-10-26
Recall event number
72013

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗